Model
|
T2RC
|
T2C
|
Workspace
|
W
|
16(406)
|
16(406)
|
D
|
13(330)
|
13(330)
|
H
|
15(381)
|
15(381)
|
Exterior
|
W
|
26(660)
|
27(686)
|
D
|
26(660)
|
27(686)
|
H
|
67(1702)
|
65(1651)
|
Temperature
|
T2RC
|
T2C
|
Low
|
C
|
-75
|
-75
|
High
|
C
|
+200
|
+200
|
Humidity capability
|
20% to 95% RH in the dry bulb range of +20°C to 85°C as limited by a 3°C dewpoint.
|
Note: Test data based on 24°C ambient, sea level, 60Hz. On 50 Hz or higher than 24°C ambient, performance may be reduced. Consult factory regarding any special cooling requirements. CFC-free refrigerants are used exclusively on all Tenney chambers.
Change Rates in Minutes
|
Ambient to C Chamber Empty
|
200°C
|
90 Minutes
|
150°C
|
40 Minutes
|
100°C
|
25 Minutes
|
93°C
|
17 Minutes
|
65°C
|
9 Minutes
|
2°C
|
5 Minutes
|
-7°C
|
7 Minutes
|
-12°C
|
10 Minutes
|
-40°C
|
30 Minutes
|
-54°C
|
45 Minutes
|
-65°C
|
70 Minutes
|
-73°C
|
100 Minutes
|
-75°C
|
Ultimate
|
Utilities, etc
|
Refrigeration
|
(2) 1/2) HP+ 1/2 HP
|
(2) 1/2) HP + 1/2 HP
|
Heater Capacity
|
500W
|
500W
|
Humidifier
|
W
|
500
|
—
|
GPH
|
0.15
|
—
|
AMPS @ 115V. 1Ph
|
18
|
18
|
AMPS Fuse
|
20
|
20
|
Unit Weight
|
LBS/KG
|
350/158
|
350/158
|