Facility Requirements
|
Power
|
60Hz only, system does not operate at 50Hz
115 ±10% VAC, 60Hz, 20amp (15amp available)
220 ±10% VAC, 60Hz, 20amp
|
Compressed Air
|
Clean, Dry Air (CDA)
|
Filtered to 5 micron particulate contamination
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.
Dew point: <-70°C @ 6.2 BAR (90PSI)
|
Air Supply Pressure
|
5.5 to 7.6 BAR (90 to 110 PSIG)
|
Total Air Flow Rate Required
|
5.7 l/s (12 scfm)
2.8 l/s (6scfm) minimum
|
Air Supply Temperature
|
+20° to +25°C; +22°C nominal
|
Operating Environment
|
Operating Temperature
|
+20° to +28°C; +23°C nominal
|
Humidity
|
0 to 60%; 45% nominal
|
Service and Safety
|
Refrigerants
|
HCFC and CFC-free, non-toxic, non-flammable
|
Serviceability
|
Auto-diagnostics and field replaceable modules
|
Over Temperature Protection
|
+230°C (factory set): Operator can set high and low air temperature limits
|
Weight and Dimensions
|
Dimensions
|
Controller
|
26.7cm (10.5 inches) wide X 20.7cm (8.13 inches) high X 9.8cm (3.8 inches) deep
|
Chiller
|
51.3cm (20.2 inches) wide X 41.4cm (16.3 inches) high X 74.9cm (29.5 inches) deep
|
Weight
|
System
|
112.0kg (247lbs.)
|
Chiller
|
74.5kg (157lbs)
|
Chiller with Cart
|
100.8kg (215lbs)
|
Noise Level
|
<60dBA
|