Frequency range |
20 Hz to 3.6 GHz |
Reference frequency |
aging: 1 × 10-7/year, optionally 2 × 10-8/year |
Absolute power measurement |
using R&S®NRP-Zxx power sensor or external power meters |
Uncertainty |
with R&S®NRP-Z27/-Z37: 0.083 dB (up to 4.2 GHz, +15 °C to +35 °C) |
Relative level measurement |
Measurement range |
+30 dBm to -140 dBm, depending on frequency |
Linearity
(+20 dBm to -140 dBm) |
±0.01 dB ± 0.005 dB per 10 dB step |
AM modulation measurements |
Modulation depth |
0 % to 100 % |
Measurement uncertainty |
1 % of reading, 50 Hz to 50 kHz;
0.4 % of reading, 90 Hz to 150 Hz, in range from 5 % to 99 % |
Modulation frequency |
10 Hz to 100 kHz |
Inherent distortion |
0.3 % |
FM modulation measurements |
Frequency deviation |
max. 500 kHz |
Measurement uncertainty |
1 % of reading, 50 Hz to 100 kHz |
Modulation frequency |
10 Hz to 200 kHz |
Inherent distortion |
0.1 % |
φM modulation measurements |
Phase deviation |
max. 10 000 rad |
Measurement uncertainty |
1 % of reading, 50 Hz to 100 kHz |
Modulation frequency |
10 Hz to 100 kHz |
Inherent distortion |
0.1 % |
Audio measurements |
Frequency range |
DC, 20 Hz to 1 MHz |
Level ranges |
0.4 V, 4 V |
Measurement uncertainty |
1 % of reading, 20 Hz to 100 kHz |
Spectrum analyzer |
Frequency range |
20 Hz to 3.6 GHz |
Resolution bandwidths |
10 Hz to 50 MHz; FFT filters: 1 Hz to 30 kHz; channel filters; EMI bandwidths |
Video bandwidths |
1 Hz to 10 MHz |
Displayed average noise level (RBW 1 Hz) |
1 GHz |
typ. -158 dBm |
26 GHz |
- |
43 GHz |
- |
50 GHz |
- |
Trace detectors |
Max. peak, min. peak, auto peak, sample, RMS, average, quasi-peak |
Phase noise |
typ. -133 dBc (1 Hz) at 10 kHz from carrier, 640 MHz carrier frequency |
Sweep time |
Span > 10 Hz |
2.5 ms to 16 000 s |
Span 0 Hz (zero span) |
1 μs to 16 000 s |
Image frequency rejection |
f < 3.6 GHz |
typ. 110 dB |
f > 3.6 GHz |
- |
f > 3.6 GHz,
with R&S®FSMR-B2 option |
- |
Display |
21 cm LC TFT color display (8.4"") |
Resolution |
800 × 600 pixel (SVGA resolution) |
Pixel failure rate |
< 1 × 10-5 |
Mass memory |
hard disk,
USB flash disk (not supplied) |
Data storage |
> 500 instrument settings and traces |
Temperature |
Operating temperature range |
+5 °C to +40 °C |
Permissible temperature range |
0 °C to +50 °C |
Storage temperature range |
-40 °C to +70 °C |
Climatic loading |
+40 °C at 95 % relative humidity
in line with EN 60068-2-30 |
Power Supply |
AC supply |
100 V to 240 V, 3.1 A to 1.3 A;
50 Hz to 400 Hz, class of protection I in
line with VDE 411 |
Power consumption |
typ. 130 VA |
Safety |
in line with EN 61010-1, IEC 61010-1,
UL 61010-1,
CAN/CSA-C22.2 No. 61010-1-4 |
Test mark |
VDE, GS, CSA, CSA-NRTL |
Dimensions (W × H × D) |
435 mm × 192 mm × 460 mm
(17.13 in × 7.56 in × 18.11 in) |
Weight, without options, nominal |
14.6 kg (32.19 lb) |