| TECHNICAL DETAILS | 
| Output Voltage | 0 V - 360 VAC (RMS)
0V - +/-500 VDC | 
| Output Current | 26 A (RMS) continuous
47 A (RMS) short-term (max. 3 s)
200 A repetitive peak | 
| SPECIFICATIONS | 
| Output frequency | Output frequency DC - 5,000Hz | 
| Frequency accuracy, stability | 100ppm | 
| DC offset with AC signal | <20mV with linear load | 
| Slew rate | 8V/us | 
| REGULATION | 
| Voltage sense | Internal or external, 4 wires | 
| Distortion (THD) | Less than 0.5%, @50/60Hz | 
| Output voltage stability | Better than 0.1% | 
| Output voltage accuracy | Better than 0.5% | 
| Max. compensable drop on wire | 5% f.s. | 
| Current limiter | 2A to Imax for f < 75Hz (NetWave 3)
5A to Imax for f < 75Hz (NetWave 7.x) | 
| Protection | Over current, over voltage,
over temperature, low voltage | 
| OUTPUTS | 
| DUT connection | 4mm safety lab connectors
DUT adapter with connector
(depends on country of use) | 
| DISPLAY AND CONTROLS | 
| Display | 2-Line LCD, 40 characters | 
| LED indicators | Power On
Active output channel
Trigger
Functional status hard disk | 
| Operation | 6 function keys,
Test On key: ON/OFF key for the
power source | 
| INTERFACES | GPIB, Ethernet
USB (for memory stick)
RS 232 (input from DPA analyzer)
Frame bus (internal system bus) |