V/A/Hz
|
Usable Bandwidth:
|
10 Hz..3.5 kHz
|
Harmonics on 400Hz fundamental:
|
9th
|
Typical accuracy for 400Hz fundamental:
|
5%
|
|
Power, battery life
|
Usable Bandwidth:
|
20 Hz..2 kHz
|
Harmonics on 400Hz fundamental:
|
5th
|
Typical accuracy for 400Hz fundamental:
|
10%
|
|
Harmonics
|
Usable Bandwidth:
|
10 Hz..3.5 kHz
|
Harmonics on 400Hz fundamental:
|
9th
|
Typical accuracy for 400Hz fundamental:
|
10% channel 1, 50% channel 2
|
|
Note
|
|
Accuracies are stated as ± (percentage of reading + counts) without probes unless otherwise noted.
|
|
Specifications below are valid for signals with a fundamental between 40 and 70 Hz.
|
|
Input Characteristics
|
Input impedance
|
1 MΩ, 20 pF
|
Voltage rating
|
600V rms, CAT III
|
|
V/A/Hz Display
|
True-rms voltage (ac + dc)
|
|
Ranges:
|
5.000 V, 50.00 V, 500.0 V, 1250 V*
|
Accuracy:
|
±(1% + 10 counts)
|
|
|
True-rms current (ac + dc)
|
|
Ranges:
|
50.00 A, 500.0 A, 5.000 kA, 50.00 kA, 1250 kA
|
Accuracy:
|
±(1% + 10 counts)
|
|
|
Frequency
|
|
Ranges:
|
40.0 to 15.9 kHz
|
Accuracy:
|
±(0.5% + 2 counts)
|
|
|
CF Crest factor
|
|
Ranges:
|
1.0 - 10.0
|
Accuracy:
|
±(0.5% + 1 count)
|
|
Power Display
|
Watts, VA, VAR
|
|
|
1-phase and 3-phase,3 conductor balanced loads
|
Ranges:
|
250 W - 1.56 GW
|
Accuracy:
|
±(4% + 4 counts) Fundamental Power
|
Accuracy:
|
± (2 % + 6 counts) Total Power
|
|
|
Power Factor
|
|
Range:
|
0 - 1.0
|
Accuracy:
|
±0.04
|
Displacement Power Factor, Cos .F
|
|
Range:
|
0.25 - 0.9
|
Accuracy:
|
±0.04
|
Range:
|
0.90 - 1.0
|
Accuracy:
|
±0.03
|
|
|
Frequency Fundamental
|
|
Ranges:
|
40.0 to 70.0 Hz
|
Accuracy:
|
±(0.5% + 2 counts)
|
|
Harmonics Display
|
Voltage, Current, Frequency
|
|
Ranges:
|
Fundamental to 51st harmonic
|
Accuracy:
|
|
Fundamental:
|
VA ±(3% + 2 counts) W ±(5% + 2 counts)
|
2 to 31st harmonic:
|
VA ±(5% +3 counts) W ±(10% +10 counts)
|
32 to 51st harmonic:
|
VA ±(15% +5 counts) W ±(30% +5 counts)
|
|
|
Frequency Fundamental
|
|
Ranges:
|
40 Hz to 70 Hz
|
Accuracy:
|
±0.25 Hz
|
|
|
Phase
|
|
Range:
|
V, A (between Fundamental & Harmonics)
|
Accuracy:
|
±3º to ±15º
|
Range:
|
W (between Voltage Fundamental & Current Harmonics)
|
Accuracy:
|
±5º to ±15º
|
|
|
K-factor (Current and Power)
|
|
Range:
|
1.0 to 30.0
|
Accuracy:
|
±10%
|
|
|
THD Total Harmonic Distortion
|
|
Range:
|
0.00 - 99.99
|
Accuracy:
|
±(3% + 8 counts)
|
|
Sags and Swells
|
Recording times:
|
4 min to 16 days (selectable)
|
Vrms Actual, Vrms max, min(AC + DC)
|
|
Ranges:
|
5.000V, 50.00V, 500.0V, 1250V*
|
Accuracy:
|
Readings ±(2% +10 counts); Cursor readings ± (2% + 12 counts)
|
|
|
Arms Actual, Arms max, min (AC + DC)
|
|
Ranges:
|
50.00A, 500.0A, 5.000 kA, 50.00 kA
|
Accuracy:
|
±(2% +10 counts)
|
|
Transient Capture
|
Minimum pulse width:
|
40 ns
|
Useful bandwidth input 1:
|
DC to 1 MHz
|
Number of transients:
|
40
|
Voltage threshold settings:
|
20%, 50%, 100%, 200% above or below reference
|
|
R, C, Diode, Continuity
|
Resistance ranges:
|
500.0 Ω, 5.000 kΩ, 50.00 kΩ, 500.0 kΩ, 5.000 MΩ, 30.00 MΩ
|
Resistance accuracy:
|
±(0.6% + 5 counts)
|
Capacitance ranges:
|
50.00 nF, 500.0 nF, 5.000 µF, 50.00 µF, 500.0 µF
|
Capacitance accuracy:
|
±(2% + 10 counts)
|
Diode voltage:
|
Accuracy ±(2% + 5 counts)
|
Continuity:
|
Beeper on at < 30 Ω ± 5 Ω
|
Max current:
|
0.5 mA
|
|
Inrush Current
|
Inrush times:
|
1 s, 5 s, 10 s, 50 s, 100 s, 5 min
|
Current ranges:
|
1 A, 5 A, 10 A, 50 A, 100 A, 500 A, 1000 A
|
Accuracy:
|
±5% of full scale
|
|
Temperature (with accessory)
|
Range:
|
-100 °C - 400 °C
|
Accuracy:
|
±(0.5% + 5 counts)
|
|
Scope Display
|
Measurements:
|
dc, ac, ac+dc, peak, peak-peak, frequency, duty cycle, phase, pulse width, crest factor
|
Time ranges:
|
20 ns/div to 60 s/div
|
Max sampling rate:
|
25 MS/s
|
Bandwidth
|
|
Voltage channel [1]:
|
20 MHz at inputs, 20 MHz with VPS40 probe (included with Fluke43B), 1 MHz with TL24 Leads
|
Current channel [2]:
|
15 kHz at inputs, 10 kHz with included clamps
|
Coupling:
|
AC, DC (10 Hz - 3 dB)
|
Vertical sensitivity:
|
5 mV/div to 500V/div
|
Vertical resolution:
|
8 bit (256 levels)
|
Record length:
|
512 samples per channel
|
Timebase modes:
|
Normal, roll, single
|
Pre-trigger:
|
Up to 10 divisions
|
Trigger Source:
|
Input 1 or Input 2 or automatic selection
|
Trigger Mode:
|
Automatic Connect-and-View™, Free Run, and Single Shot
|
|
|
Connect-and-View™:
|
Advanced automatic triggering that recognizes signal patterns
|
|
Automatically adjusts triggering, timebase and amplitude and displays stable pictures
|
|
Memories
|
|
20 (screens, settings, data)
|
|
Recording
|
Recording times:
|
4 min to 16 days (selectable)
|
Parameters:
|
Choose one or two parameters from one of the groups below:
|
|
Volts/Ampères/Hertz
|
|
Watts, VA, VAR, PF, DPF, Frequency
|
|
Harmonics, THD, Volts (Fund. & Harmonic), Ampères (F&H) Watts(F&H)
|
|
Frequency (H), %(H) of total, Phase(H), KF
|
|
Temperature
|
|
Resistance, Diode, Continuity, Capacitance
|
|
Scope: DC Voltage, DC Current, AC Voltage, AC Current, Frequency,
|
|
Pulse Width + or -, Phase, Duty cycle + or -, Peak max, Peak min, Peak min-max, Crest Factor
|
|
Note
|
|
*Rated EN 61010-1 600 V CAT II CSA
|
|