Frequency range |
2 Hz to 44 GHz |
Resolution bandwidths |
Standard filters (-3 dB) |
1 Hz to 10 MHz |
EMI filters (-6 dB) |
1 Hz, 10 Hz, 100 Hz, 200 Hz, 1 kHz, 9 kHz,
10 kHz, 100 kHz, 120 kHz, 1 MHz, 10 MHz |
Detectors (receiver mode) |
max. peak, min. peak, quasi-peak, RMS, average, average with meter time constants (CISPR-average), RMS-average (CISPR RMS) |
APD measurement function |
Minimum amplitude probability |
10-7 |
Maximum acquisition time |
120 s |
Acquisition bandwidth (-6 dB) |
200 Hz, 9 kHz, 120 kHz, 1 MHz |
Third-order intercept (TOI) |
R&S®ESW8, R&S®ESW26, R&S®ESW44,
RF attenuation = 0 dB, level 2 × -15 dBm,
Δf > 5 × RBW, preselection off, preamplifier off |
10 MHz ‰¤ fin < 1 GHz, > +20 dBm, typ. 25 dBm |
1 dB compression |
RF attenuation = 0 dB,
preselection off, preamplifier off |
fin ‰¤ 3 GHz, +15 dBm (nom.) |
Preselection filters |
Receiver mode status |
always on |
Analyzer mode status |
on/off (selectable) |
Number of preselection filters |
21 |
Preamplifier (switchable) (1 kHz to 8 GHz) |
20 dB (nom.) |
Displayed average noise level (DANL) |
R&S®ESW8 (receiver mode, RF attenuation = 0 dB, termination = 50 Ω, logarithmic scaling normalized to 1 Hz RBW, RBW = 1 kHz, VBW = 1 Hz, +5 °C to +40 °C, preselection on, preamplifier on) |
Receiver mode |
Time domain scan |
CISPR band B, 150 kHz to 30 MHz, RBW = 9 kHz,
measurement time = 100 ms, peak detector |
110 ms (total measurement time) |
CISPR band B, 150 kHz to 30 MHz,
RBW = 9 kHz, measurement time = 1 s,
quasi-peak detector and CISPR-average detector |
2 s (total measurement time) |
CISPR band C/D, 30 MHz to 1000 MHz,
RBW = 120 kHz, measurement time = 10 ms, peak detector |
620 ms (total measurement time) |
CISPR band C/D, 30 MHz to 1000 MHz,
RBW = 9 kHz, measurement time = 10 ms, peak detector |
840 ms (total measurement time) |
CISPR band C/D, 30 MHz to 1000 MHz,
RBW = 120 kHz, measurement time = 1 s, quasi-peak detector |
80 s (total measurement time) |
CISPR band C/D, 30 MHz to 1000 MHz,
RBW = 9 kHz, measurement time = 1 s, quasi-peak detector and CISPR-average detector |
67 s (total measurement time) |
Total measurement uncertainty |
Preselection off (10 MHz < f ‰¤ 3.6 GHz) |
±0.27 dB |
Dimensions (nom.)
(W × H × D, including front handles and rear feet) |
462 mm × 240 mm × 504 mm
(18.15 in × 9.44 in × 19.81 in) |
Net weight, without options (nom.) |
25.2 kg (55.56 lb) |
Display |
30.7 cm (12.1"") WXGA color touchscreen |
Resolution |
1280 × 800 pixel (WXGA resolution) |
Data storage |
Internal |
solid state disk ‰¥ 32 Gbyte |
External |
supports USB 2.0 compatible memory
devices |
Power supply |
AC input voltage range |
100 V to 240 V |
AC supply frequency |
50 Hz to 60 Hz/400 Hz |
Max. input current |
7.3 A (100 V) to 4.6 A (240 V) |
Power consumption |
R&S®ESW44 without options |
200 W |
R&S®ESW44 with all options |
300 W (meas.) |