- High or low temperature drying
- Stainless steel interior
- Painted, cold-rolled steel pressure vessel
- Galvanized steel exterior with powder coating
- Electric plate heating inside pressure vessel
- Double rotary vane vacuum pumps
- Large and insulated tempered glass window
- Two shelves
NOTE: Not designed for volatile, explosive, or battery applications. Please consult factory for these applications.
Performance
- Temperature range: Ambient to +200°C
- Control tolerance: <1°C
- Heating rate: >2°C/minute
- Vacuum: 8mm/hg
Controller – Watlow F4
- Single Channel
- Profiling: Forty Profiles, 256 Steps
- Automatic & Manual Control
- 1 Input: Universal Type (RTD is Std.)
- 4 Digital Inputs
- 2 Auxiliary Analog Inputs - Optional
- 2 Control Outputs: Time Proportioned, On/Off, or 4-20 ma
- 2 Alarm Outputs
- 8 Digital Event Outputs
- 2 Retransmit Outputs - Optional
- Communications: EIA-232 and EIA-485 with Modbus RTU protocol
Controller – Yokogawa UT150L
- Non-profiling type controller
- 1/16 DIN
- Single channel
- Automatic control only
- 1 process input: Type J Thermocouple
- 1 heat control output and 1 alarm output
- No communications