System |
Control method |
DC Proportional, Integral, Derivative (PID) |
Remote control |
RS232, IEEE-488 |
Local control |
Touch pad, vacuum fluorescent display |
Temperature accuracy |
±0.5°C ( when calibrated against NIST standard) |
Temperature stability |
±0.1°C |
Temperature resolution |
0.1°C |
Ambient operating temperature |
+10° to +30°C |
Humidity operating range |
0 to 90%, non-condensing |
Overheat protection temperature |
Redundant RTD |
Temperature Display |
0.1°C resolution |
Audible Noise of System |
62 dbA |
Refrigerants |
HCFC-free and CFC-free, non-toxic, non-flammable |
Industry Certifications |
Compliant with CE and ETL and conforms to the SEMI S2-93A safety |
Facility Requirements |
Power |
200 - 250 VAC (230V nominal), 60Hz, 30 amp, 1phase |
Clean, Dry Air (CDA) |
Filtered to 5 micron particulate contamination.
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.
Dew point: <-10°C @ 6.2 BAR (90PSI) |
Fluid |
1.5 liters (50.7 fl. oz.). For fluid type, contact factory |
Vacuum |
600 mm HG. |
WEIGHT |
Complete System (standard ), packed with accessories |
TP03500A |
317 kg. (700 lb.) approximately |
TP03500C |
Check with factory |
TP03500D |
Check with factory |
TP03500E |
Check with factory |