Facility Requirements |
Power |
60Hz only, system does not operate at 50Hz
200 - 250 VAC (230V nominal), 60Hz, 30 amp, 1phase |
Compressed Air |
Clean, Dry Air (CDA) |
Filtered to 5 micron particulate contamination.
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.
Dew point: <-10°C @ 6.2 BAR (90PSI) |
Air Supply Pressure |
6.2 to 7.6 BAR (90 to 110 PSIG) |
Total Air Flow Rate Required |
7.1 to 14.2 l/s (15-30 scfm)
11.8 l/s (25 scfm) nominal |
Air Supply Temperature |
+20° to +25°C; +22°C nominal |
Operating Environment |
Operating Temperature |
+20° to +28°C; +23°C nominal |
Humidity |
0 to 60%; 45% nominal |
Service and Safety |
Refrigerants |
HCFC and CFC-free, non-toxic, non-flammable |
Serviceability |
Auto-diagnostics and field replaceable modules |
Over Temperature Protection |
+230°C (factory set): Operator can set high and low air temperature limits |
Weight and Dimensions |
Dimensions |
Base |
Width: 61.0 cm (24 in.), Depth: 72.4 cm (28.5 in.)
Height: 108 cm (42.5 in.) |
System Weight |
Not packed: 236 kg (520 lbs.)
Packed: 365 kg (805 lbs.) |
Mobility |
Four static dissipative, swivel caster wheels |
Max. Reach |
160.0cm (63 in.) |
Max. Operating Height |
130.3 cm (51.2 in.) - extended height option: 188.0 (74.0 in.) |
Min. Operating Height |
69.1 cm (27.2 in.) - extended height option: 81.3 (32.0 in.) |
Noise Level |
<65dBA |