Facility Requirements
|
Power
|
200 - 250 VAC (230V nominal),
50/60Hz 10 amp, 1phase
|
Compressed Air
|
Clean, Dry Air (CDA)
|
Filtered to 5 micron particulate contamination.
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant.
Dew point: ambient to -40°C
|
Air Supply Pressure
|
5.6 to 7kg /cm2 (80 to 90 PSIG)
5.6 kg/cm (90 PSIG) nominal
|
Total Air Flow Rate Required
|
4 to 5 l/s (8-10 scfm)
4.5 l/s (9 scfm) nominal
|
Air Supply Temperature
|
+20° to +28°C; +23°C nominal
|
Operating Environment
|
Operating Temperature
|
+20° to +28°C; +23°C nominal
|
Humidity
|
0 to 60%; 45% nominal
|
Service and Safety
|
Refrigerants
|
HCFC and CFC-free, non-toxic, non-flammable
|
Serviceability
|
Auto-diagnostics and field replaceable modules
|
Over Temperature Protection
|
+230°C (factory set): Operator can set high and low air temperature limits
|
Weight and Dimensions
|
Dimensions
|
Controller
|
40.1cm (15.8 inches) WIDE x 14.8cm (5.8 inches) HIGH x 48.9cm (19.2 inches) DEEP
|
Thermal Wand
|
8cm (3 inches) DIAMETER, 36cm (14 inches) LONG
|
Stand
|
48.25cm (19 inches) approximately above benchtop to top of post
|
Reach from Vertical Post
|
standard, 51cm (20 inches)
|
Weight
|
Controller with Thermal Wand and Hose: 14.5 kg (32lbs.)
Thermal Wand Stand Assembly: 7kg (5.5 lbs.)
Thermal Cap: .45 kg (1lb.)
|
Noise Level
|
<70dBA
|