 Facility Requirements 
 | 
| 
 Power 
 | 
 60Hz only, system does not operate at 50Hz 
115 ±10% VAC, 60 Hz 15 amp, 1phase 
230 ±10% VAC, 60 Hz 10 amp, 1phase 
 | 
| 
 Compressed Air 
 | 
| 
 Clean, Dry Air (CDA) 
 | 
 Filtered to 5 micron particulate contamination. 
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant. 
Dew point: ambient to -40°C 
 | 
| 
 Air Supply Pressure 
 | 
 5.6 to 7kg /cm2 (80 to 90 PSIG) 
5.6 kg/cm (90 PSIG) nominal 
 | 
| 
 Total Air Flow Rate Required 
 | 
 4 to 5 l/s (8-10 scfm) 
4.5 l/s (9 scfm) nominal 
 | 
| 
 Air Supply Temperature 
 | 
 +20° to +28°C; +23°C nominal 
 | 
| 
 Operating Environment 
 | 
| 
 Operating Temperature 
 | 
 +20° to +28°C; +23°C nominal 
 | 
| 
 Humidity 
 | 
 0 to 60%; 45% nominal 
 | 
| 
 Service and Safety 
 | 
| 
 Refrigerants 
 | 
 HCFC and CFC-free, non-toxic, non-flammable 
 | 
| 
 Serviceability 
 | 
 Auto-diagnostics and field replaceable modules 
 | 
| 
 Over Temperature Protection 
 | 
 +230°C (factory set): Operator can set high and low air temperature limits 
 | 
| 
 Weight and Dimensions 
 | 
| 
 Dimensions 
 | 
 Controller 
 | 
 40.1cm (15.8 inches) WIDE x 14.8cm (5.8 inches) HIGH x 48.9cm (19.2 inches) DEEP 
 | 
| 
 Thermal Wand 
 | 
 8cm (3 inches) DIAMETER, 36cm (14 inches) LONG 
 | 
| 
 Stand 
 | 
 48.25cm (19 inches) approximately above benchtop to top of post 
 | 
| 
 Reach from Vertical Post 
 | 
 standard, 51cm (20 inches) 
 | 
| 
 Weight 
 | 
 Controller with Thermal Wand and Hose: 14.5 kg (32lbs.) 
Thermal Wand Stand Assembly: 7kg (5.5 lbs.) 
Thermal Cap: .45 kg (1lb.) 
 | 
| 
 Noise Level 
 | 
 <70dBA 
 |