 Facility Requirements 
 | 
| 
 Power 
 | 
 60Hz only, system does not operate at 50Hz 
115 ±10% VAC, 60Hz, 20amp 
220 ±10% VAC, 60Hz, 20amp 
 | 
| 
 Compressed Air 
 | 
| 
 Clean, Dry Air (CDA) 
 | 
 Filtered to 5 micron particulate contamination. 
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant. 
Dew point: <-70°C @ 6.2 BAR (90PSI) 
 | 
| 
 Air Supply Pressure 
 | 
 5.5 to 7.6 BAR (90 to 110 PSIG) 
 | 
| 
 Total Air Flow Rate Required 
 | 
 5.7 l/s (12 scfm) 
2.8 l/s (6 scfm) minimum 
 | 
| 
 Air Supply Temperature 
 | 
 +20° to +25°C; +22°C nominal 
 | 
| 
 Operating Environment 
 | 
| 
 Operating Temperature 
 | 
 +20° to +28°C; +23°C nominal 
 | 
| 
 Humidity 
 | 
 0 to 60%; 45% nominal 
 | 
| 
 Service and Safety 
 | 
| 
 Refrigerants 
 | 
 HCFC and CFC-free, non-toxic, non-flammable 
 | 
| 
 Serviceability 
 | 
 Auto-diagnostics and field replaceable modules 
 | 
| 
 Over Temperature Protection 
 | 
 +230°C (factory set): Operator can set high and low air temperature limits 
 | 
| 
 Weight and Dimensions 
 | 
| 
 Dimensions 
 | 
 Controller 
 | 
 26.7cm (10.5 inches) wide X 20.7cm (8.13 inches) high X 9.8cm (3.8 inches) deep 
 | 
| 
 Chiller 
 | 
 51.3cm (20.2 inches) wide X 41.4cm (16.3 inches) high X 74.9cm (29.5 inches) deep 
 | 
| 
 Weight 
 | 
 System 
 | 
 112.0kg (247lbs.) 
 | 
| 
 Chiller 
 | 
 74.5kg (157lbs) 
 | 
| 
 Chiller with Cart 
 | 
 100.8kg (215lbs) 
 | 
| 
 Noise Level 
 | 
 <60dBA 
 |