 Facility Requirements 
 | 
| 
 Power 
 | 
 200 - 250 VAC (230V nominal), 50/60Hz, 30 amp, 1phase 
 | 
| 
 Compressed Air 
 | 
| 
 Clean, Dry Air (CDA) 
 | 
 Filtered to 5 micron particulate contamination. 
Oil Content: <0.1 ppm, by weight, filtered to 0.01 micron oil contaminant. 
Dew point: <-10°C @ 6.2 BAR (90PSI) 
 | 
| 
 Air Supply Pressure 
 | 
 6.2  to 7.6 BAR (90 to 110 PSIG) 
 | 
| 
 Total Air Flow Rate Required 
 | 
 7.1 to 14.2 l/s (15-30 scfm) 
11.8 l/s (25 scfm) nominal 
 | 
| 
 Air Supply Temperature 
 | 
 +20° to +25°C; +22°C nominal 
 | 
| 
 Operating Environment 
 | 
| 
 Operating Temperature 
 | 
 +20° to +28°C; +23°C nominal 
 | 
| 
 Humidity 
 | 
 0 to 60%; 45% nominal 
 | 
| 
 Service and Safety 
 | 
| 
 Refrigerants 
 | 
 HCFC and CFC-free, non-toxic, non-flammable 
 | 
| 
 Serviceability 
 | 
 Auto-diagnostics and field replaceable modules 
 | 
| 
 Over Temperature Protection 
 | 
 +230°C (factory set): Operator can set high and low air temperature limits 
 | 
| 
 Weight and Dimensions 
 | 
| 
 Dimensions 
 | 
 Base 
 | 
 Width: 61.0 cm (24 in.), Depth: 72.4 cm (28.5 in.) 
Height: 108 cm (42.5 in.) 
 | 
| 
 System Weight 
 | 
 Not packed: 236 kg (520 lbs.) 
Packed: 365 kg (805 lbs.) 
 | 
| 
 Mobility 
 | 
 Four static dissipative, swivel caster wheels 
 | 
| 
 Max. Reach 
 | 
 160.0cm (63 in.) 
 | 
| 
 Max. Operating Height 
 | 
 130.3 cm (51.2 in.) - extended height option: 188.0 (74.0 in.) 
 | 
| 
 Min. Operating Height 
 | 
 69.1 cm (27.2 in.) - extended height option: 81.3 (32.0 in.) 
 | 
| 
 Noise Level 
 | 
 <65dBA 
 |