PSURGE 30.2 Mainframe |
Charging voltage range |
3...33kV |
Capacitors |
9uF / 36uF |
Stored energy |
5kJ at max. charging voltage |
HV output |
terminals inside the test cabinet |
Power mains |
208, 230 or 240V
50/60Hz, 1800VA |
HV switch |
triggered spark gap |
Impulse polarity |
positive, negative, alternating |
Impulse circuit |
floating, max. voltage between common and ground 500V AC |
External synchronization |
0°....360° in 1° steps |
Dimensions (W x D x H) |
PSURGE 30.2 |
60 x 100 x 190 cm |
Test cabinet |
39 x 29 x 32 cm |
Weight |
approx. 320 kg |
Impulse Module PS30-CW |
Impulse type |
Combination wave |
Voltage front time |
1.2us ±30% |
Voltage decay time |
50us ±20% |
Voltage undershoot |
‰¤30% |
Voltage amplitude range |
3...30kV |
Impulse repetition |
Up to 6 impulses per minute at max. voltage |
Output impedance |
2 Ohm |
Current front time |
8us ±20% |
Current decay time |
20us ±20% |
Current undershoot |
‰¤30% |
Current amplitude range |
1.5...15kA |
Impulse Module PS30-8x20 |
Impulse type |
Current |
Voltage front time |
not defined |
Voltage decay time |
not defined |
Undershoot |
not defined |
Charging voltage range |
3...15kV |
Impulse repetition |
Up to 2 impulses per minute at max. current |
Output impedance |
0.5 Ohm |
Current front time |
8us ±20% |
Current decay time |
20us ±20% |
Undershoot |
‰¤20% |
Current amplitude range |
6...30kA |
Impulse Module PS30-10x350 |
Impulse type |
current |
Voltage front time |
not defined |
Voltage decay time |
not defined |
Charging voltage range |
3...15kV |
Impulse repetition |
Up to 2 impulses per minute at max. current |
Output impedance |
12 Ohm |
Current front time |
10us ±20% |
Current decay time |
350us ±20% |
Current amplitude range |
240...1200A |
Impulse Module PS30-10x1000 |
Impulse type |
Current |
Voltage front time |
not defined |
Voltage decay time |
not defined |
Charging voltage range |
3...15kV |
Impulse repetition |
Up to 2 impulses per minute at max. current |
Output impedance |
37 Ohm |
Current front time |
10us ±20% |
Current decay time |
1000us ±20% |
Current amplitude range |
80...400A |
Coupling and decoupling network FP-SURGE 3010 |
Impulse shape |
Combination wave only |
EUT line voltage |
480V AC
110V DC |
EUT current |
10A |
Imp. amplitude |
30kV / 15kA |
Coupling modes |
Line to neutral
Line or neutral to ground
Line and neutral to ground |
Coupling capacitors |
2x9uF / 1x18uF |
Residual voltage at EUT line input |
15% of the applied test voltage, limited to approx. 1200V |
Dimensions (W x D x H) |
FP-SURGE 3010 |
60 x 100 x 190 cm |
Test cabinet |
40 x 40 x 80 cm |
Weight |
FP-SURGE 3010 |
approx. 250 kg |