GENERAL DATA |
Control power |
85 V - 264 V
50/60 Hz |
User test storage |
unlimited |
Remote interface |
Ethernet, RJ45 |
Display |
7“/ 800x480 / 24 bit with touch-screen |
USB |
for USB memory stick |
AUX. interface |
D-sub 37p for external CDN, external transformer etc. |
External trigger input |
5 V TTL |
Synch input |
BNC, 10 V - 264 V AC |
Trigger output |
5 V TTL |
External start / stop input |
5 V TTL, starts / stops predefined test sequence |
EUT failed input |
5 V TTL |
Analog output |
0 - 10 V, for use with external options |
Warning lamp output |
2 x 24 V / 1 A DC |
Safety circuit |
stops the test when unlocked |
Dimensions |
19“/ 4U |
(W x H x D) |
45 x 18 x 49 cm |
Weight |
30 kg |
IEC / EN 61000-4-4 EDITION 2 EFT / BURST |
Output voltage |
0.2 - 5.0 kV ±10% at coaxial output |
Spike frequency |
1 Hz - 1 MHz |
Polarity |
pos / neg / alternate |
Burst duration |
10 us - 1 s |
Output impedance |
50 Ohms |
Burst period |
1 ms - 10 s |
Rise time |
5 ns ±30% |
Test time |
1 s- 999 minutes |
Impulse duration |
50 ns ±30% at 50 Ohm
50 ns -15 +100 ns at 1000 Ohm |
Trigger |
automatic, manual,
external trigger input |
Burst mode |
normal, continuous, real, random |
Integrated single phase coupling / decoupling network |
264 V AC / 16 A
220 V DC / 10 A |
IEC / EN 61000-4-5 EDITION 2 & 3 SURGE COMBINATION WAVE |
Output voltage |
0.2 - 7.0 kV ±10% Output current 0.1 - 3.5 kA ±10% |
Voltage rise time |
1.2 μs ±30% Current rise time 8 us ±20% |
Voltage duration |
50 μs ±20% Current duration 20 us ±20% |
Polarity |
pos / neg / alternate |
Integrated single phase CDN |
264 V AC / 16 A
220 V DC / 10 A |
Output impedance |
2 Ohms |
Phase synch |
0 - 359° with 1° steps or asynchronous mode |
Impulse trigger |
automatic 2 s - 100 min
manual external trigger input |
Counter preselect counter |
1 - 1000 / infinite |
Counter |
100000 |
Peak voltage monitor |
BNC output: 1000:1
display: 3 digits |
Peak current monitor |
BNC output: 1 kA/V
display: 3 digits |
IEC / EN 61000-4-11 EDITION 2 AND IEC / EN 61000-4-29 DIPS & INTERRUPTS |
Max. voltage |
264 V AC/DC |
Interrupt time
|
0.5 period - 800 periods
100 μs - 1000 minutes |
Max. current |
16 A AC/DC continuous
20 A for 5 s
23 A for 3 s
40 A for 3 s
> 500 A inrush current |
Interval time |
1 period - 800 periods synch
100 μs - 1000 minutes asynch |
Trigger |
Automatic
Manual external trigger input |
Test time |
1 s - 1000 minutes infinite |
Interrupt dip level |
0%
0% - 99% with external voltage source |
Phase synch |
0 - 359° 16 / 40 / 50 / 60 Hz
asynchronous mode |
RMS voltage monitor |
BNC output: 100:1
display: 4 digits |
RMS current monitor |
BNC output: 10 A/V
display: 4 digits |
IEC / EN 61000-4-12 EDITION 2 AND ANSI / IEEE C62.41 Ring Wave |
Max. voltage |
0.2 - 7.0 kV ± 10% |
Repetition rate |
Up to 30 pulses / min |
Frequency |
100 kHz |
Polarity |
Positive / negative / alternate |
Rise time OC |
5 μs |
Rise time SC |
1 μs |
Floating output |
Max. 460 V / AC |
Phase sync accuracy |
± 1° |
Impedance |
12 Ohm, 30 Ohm |
Damping rate |
0.4 < peak1/peak2 < 1.1
0.4 < peak3/peak2 < 1.1
0.4 < peak4/peak3 < 1.1 |
Peak voltage monitor |
BNC output: 1000:1
Display: 3 digits |
Peak current monitor |
BNC output: 1kA / V
Display: 3 digits |