IMPULSE MODULES |
PIM 100 Combination wave impulse module |
Built according to IEC/EN 61000-4-5 Ed. 1 & 2 and ANSI C62.41/45 |
1.2/50 μs open circuit up to 7.4 kV |
8/20 μs short circuit up to 3.7 kA |
Impulse voltage and current monitors |
± 1° phase synchronization |
Reliable semiconductor HV-switch |
Positive, negative and alternating polarity |
Up to 12 pulses per minute |
PIM 110 Ring wave impulse module |
Built according to IEC/EN 61000-4-12 and ANSI C62.41/45 |
100 kHz frequency, 0.5 μs rise time |
Imp. voltage up to 7.8 kV / 12 Ω, 30 Ω and 200 Ω |
Impulse voltage and current monitors |
± 1° phase synchronization |
Positive, negative and alternating polarity |
Up to 12 pulses per minute |
Reliable semiconductor HV-switch |
PIM 120 Telecom wave impulse module |
Built according to IEC/EN 61000-4-5 Ed. 1 & 2, ITU K-series and IEC 60950 |
10/700 μs open circuit voltage |
5/320 μs short circuit current |
Imp. voltage up to 7.4 kV / 15 Ω and 40 Ω |
Impulse voltage & current monitors |
Reliable semiconductor HV-switch |
Positive, negative and alternating polarity |
Up to 12 pulses per minute |
PCD 120 Four-wire coupling unit for telecom wave |
4 wire coupling unit for unshielded symmetrical operated lines |
Compliant with IEC 61000-4-5 (10/700 μs impulse only), ITU K.20 / K.21 / K.44 / K.45 |
Fully automated operation |
PIM 150 Oscillating wave impulse module |
Built according to IEC/EN 61000-4-12 Ed.1, IEC 61000-4-18 Ed.1, IEC 60255-22-1 and ANSI C37.90 |
100 kHz and 1 MHz burst frequencies |
75 ns rise time |
Impulse voltage up to 3.3 kV / 200 Ω |
Integrated, fully automatic CDN for three-phase AC and DC power lines |
EUT mains voltage up to 480 V / 277 V 16 A |
Impulse voltage monitor |
Capacitive coupling clamp available optionally (IP 4 A) |
Data line CDN available optionally (PCD 150) |
COUPLING / DECOUPLING NETWORKS |
PCD 100 Single phase CDN for combination & ring wave |
Built according to IEC/EN 61000-4-5 Edition 1 & 2 ANSI C62.41/45, TIA-968A (FCC part 68/47 CFR), Telcordia (Bellcore) GR 1089 CORE |
Fully automated test routines |
Coupling of up to two different wave-shapes without any rewiring |
For combination wave, ring wave and others (PIM 100, PIM 110) |
Up to 8 kV impulse voltage |
EUT mains up to 264 V / 16 A |
PCD 130 Three phase CDN for combination and ring wave |
Built according to IEC/EN 61000-4-5 Edition 1 & 2 ANSI C62.41/45, TIA-968A (FCC part 68/47 CFR), Telcordia (Bellcore) GR 1089 CORE |
Fully automated test routines |
Coupling of up to three different wave- shapes without any rewiring |
For combination wave, ring wave and others |
Up to 8 kV impulse voltage |
EUT mains voltage up to 690 V / 400 V |
EUT mains current up to 32 A per phase |
Correct phase angle synchronization for each coupling path |