Carrier frequency
|
Range
|
10 kHz to 2700 MHz
|
Resolution
|
0.01 Hz
|
Accuracy
|
Same as that of the reference oscillator
|
Internal reference oscillator*1
|
Frequency: 10 MHz
Start-up characteristics: After 30 minutes of operation: ≤1 x 10–7/day, after 60 minutes of operation: ≤5 x 10–8/day, Aging rate: After 24 hours of operation: ≤2 x 10–8/day,
Temperature characteristics: ±5 x 10–8/day, (0˚ to 50˚C)
|
External reference signal input
|
10 MHz, TTL Level, BNC connector on rear panel
|
Reference signal output
|
10 MHz, TTL Level, BNC connector on rear panel
|
Switching time
|
≤10 ms (time from last command until frequency has stabilized to within ±500 Hz of set frequency, during remote operation)
|
Output
|
Range
|
–143 to +23 dBm
|
Units
|
dBm, dBµV, V, mV, µV (Terminated and open voltages are selectable for dBµV, V, mV or µV.)
|
Resolution
|
0.1 dB
|
Frequency response
|
±0.5 dB referred to 0 dBm (<1280 MHz), ±1 dB referred to 0 dBm (≥1280 MHz)
|
Accuracy
|
Frequency Output level
|
10 kHz to <1280 MHz
|
≥1280 MHz
|
+17.1 to +23 dBm
|
–
|
–
|
+15.1 to +17 dBm
|
±1 dB
|
–
|
–122.9 to +15 dBm
|
±1 dB
|
±2 dB
|
–132.9 to –123 dBm
|
±3 dB
|
±4 dB
|
–143 to –133 dBm
|
–
|
–
|
Impedance
|
50 Ω, N-type connector
VSWR: ≤1.5 (<1280 MHz, ≤–3 dBm), ≤1.8 (≥1280 MHz, ≤–3 dBm)
|
Switching time
|
Time from last command until output level is stabilized, during remote operation:≤25 ms (at LEVEL NORMAL mode)
≤80 ms (when setting level is crossing over –59 dBm, at LEVEL NORMAL mode)
≤5 ms (at LEVEL CONTINUOUS mode)
|
Interference radiation
|
≤1 µV (Value is voltage terminated with 50 Ω load, measured 25 mm from front panel with a two-turn 25 mm diameter loop antenna.) Except sweep mode
|
Range |
10 kHz to 2700 MHz
|
Resolution
|
0.01 Hz
|
Accuracy
|
Same as that of the reference oscillator
|
Internal reference oscillator*1
|
Frequency: 10 MHz
Start-up characteristics: After 30 minutes of operation: ≤1 x 10–7/day, after 60 minutes of operation: ≤5 x 10–8/day, Aging rate: After 24 hours of operation: ≤2 x 10–8/day,
Temperature characteristics: ±5 x 10–8/day, (0˚ to 50˚C)
|
External reference signal input
|
10 MHz, TTL Level, BNC connector on rear panel
|
Reference signal output
|
10 MHz, TTL Level, BNC connector on rear panel
|
Switching time
|
≤10 ms (time from last command until frequency has stabilized to within ±500 Hz of set frequency, during remote operation)
|
Range |
–143 to +23 dBm
|
Units
|
dBm, dBµV, V, mV, µV (Terminated and open voltages are selectable for dBµV, V, mV or µV.)
|
Resolution
|
0.1 dB
|
Frequency response
|
±0.5 dB referred to 0 dBm (<1280 MHz), ±1 dB referred to 0 dBm (≥1280 MHz)
|
Accuracy
|
Frequency Output level
|
10 kHz to <1280 MHz
|
≥1280 MHz
|
+17.1 to +23 dBm
|
–
|
–
|
+15.1 to +17 dBm
|
±1 dB
|
–
|
–122.9 to +15 dBm
|
±1 dB
|
±2 dB
|
–132.9 to –123 dBm
|
±3 dB
|
±4 dB
|
–143 to –133 dBm
|
–
|
–
|
Impedance
|
50 Ω, N-type connector
VSWR: ≤1.5 (<1280 MHz, ≤–3 dBm), ≤1.8 (≥1280 MHz, ≤–3 dBm)
|
Switching time
|
Time from last command until output level is stabilized, during remote operation:≤25 ms (at LEVEL NORMAL mode)
≤80 ms (when setting level is crossing over –59 dBm, at LEVEL NORMAL mode)
≤5 ms (at LEVEL CONTINUOUS mode)
|
Interference radiation
|
≤1 µV (Value is voltage terminated with 50 Ω load, measured 25 mm from front panel with a two-turn 25 mm diameter loop antenna.) Except sweep mode
|
External modulation |
Input level: Approx. 2 Vp-p/600 Ω
Input impedance: Nominal 600 Ω
|
Distortion
|
≤1% (internal 1 kHz, 5 rad modulation)
|
|
400 Hz, 1 kHz (fixed oscillator)
0.1 Hz to 100 kHz (variable oscillator)
DC voltage signals equivalent peak values of internal modulating sine wave can be applied as a modulating signal using the SPECIAL FUNCTION.
|
Resolution
|
0.1 Hz
|
Frequency accuracy
|
100 ppm
|
Distortion
|
≤0.03% (fixed, 400 Hz and 1 kHz), ≤0.3% (variable, 20 Hz to 50 kHz)
|
Frequency memory |
1000 carrier frequencies (store/recall)
|
Function memory
|
100 panel settings (store recall)
|
Sweep mode |
Carrier frequency, output level, AF frequency
|
Sweep time
|
0.1 ms to 600 s, 0.01 ms resolution (minimum time depends on the switching time of each function.)
|
Marker
|
One movable marker
|
Sweep signal output
|
Staircase (saw-tooth waveform), Start point: 0 V, Stop point: 10 V
|
Modulation signal output |
Modulation signal is output when modulating. Output level: Approx. 2 Vp-p/600 Ω
|
Simultaneous modulation
|
Simultaneous modulation is possible in combinations shown below.
|
Relative value display
|
Carrier frequency, output level
|
Continuously variable output level mode
|
Continuously variable within a ±10 dB range of the set level Step size: 0.1 dB
|
Trigger function
|
Previously programmed operation procedure can be started by a trigger input through its input terminal (on rear panel, BNC connector, TTL level). Maximum program steps for triggered operation: 99 steps
|
Memory backup
|
Last settings are stored when power is turned off.
|
GPIB
|
Interface function: SH1, AH1, T5, L3, TE0, LE0, SR1, RL1, PP0, DC1, DT1, C0
|
Reverse power
|
Maximum reverse input power: 50 W (<1000 MHz), 25 W (≥1000 MHz), ±DC 50 V
|
Operating temperature
|
0˚ to 50˚C
|
Power
|
*7Vac +10%, 48 to 63 Hz, ≤270 VA
–15
|
Dimensions and mass
|
426 (W) x 177 (H) x 451 (D) mm, ≤32 kg
|
EMC*8
|
EN55011: 1991, Group 1, Class A
EN50082-1: 1992
|
Safety
|
EN61010-1:1993 (Installation Category II (Pollution Degree II)
|