STANDARDS
|
Conforms with the following:
|
o IEC 60651 Type 1, 1979 & Amendment 1 1993 & Amendment 2 2000
|
o IEC/EN 61672 – Draft March 1998 Class 1
|
o EN60651 Type 1
|
o EN60804 Type 1
|
o ANSI S1.4 –1983 Type S1
|
o ANSI S1.43 – 1997 Type 1
|
SUPPLIED MICROPHONE
|
Type 4188 Prepolarized Free-field 1/2"" Condenser Microphone
|
Nominal Sensitivity: –30 dB re 1 V/Pa or 31.6 mV/Pa
|
Frequency Range: 8Hz–16 kHz }2dB
|
Capacitance: 12 pF
|
MICROPHONE PREAMPLIFIER
|
ZC 0030
|
Extension Cables: Available in lengths of 3m and 10 m. No recalibration is required
|
MEASURING RANGES
|
Dynamic Range: 80 dB, adjustable to give full-scale readings from 80 to 140 dB in 10 dB steps
|
Max. Peak Level: 3 dB above full scale reading
|
Upper Limit (RMS) for Crest Factor = 10: 17 dB below full scale reading
|
Pulse Range: 83 dB
|
Linear operating ranges (broadband):
|
For the individual level ranges, at 1 kHz, the nominal upper boundary level minus the lowest sound pressure level measurable with a noise margin of 5 dB.
|
DISPLAY
|
128 pixel × 64 pixel dot matrix display with backlight
|
AUTO-START
|
Mediator supports a total of four timers which allow setup of measurement start times up to a month in advance
|
CALIBRATION
|
Semi-automatic, using Sound Level Calibrator Type 4231 or Multifunction Acoustic Calibrator Type 4226.
|
The initial factory calibration (sensitivity and microphone serial number) is stored for comparison with later calibrations. When using the supplied microphone the maximum allowed deviation from the initial sensitivity is }1.5 dB. An unspecified microphone can be chosen during calibration, in which case calibration can be made with practically any sensitivity.
|
Stored Calibration History: 20 latest calibrations plus initial calibration
|
MEMORY
|
2Mbytes. Up to 511 measurements can be stored by each loaded software module, including time stamp, complete setup and calibration data
|
SERIAL PRINTER
|
Measurement data can be printed on Portable Printer Type 2322 or on an IBM® Proprinter-compatible printer
|
CLOCK
|
Real-time (calendar)
|
SERIAL INTERFACE
|
Conforms to EIA/TIA 574 (RS–232), coupled as Data Terminal
|
Equipment (DTE). Cable (AO 1442) is included
|
Connector: 9–pin D-type male
|
Baud Rates: 4800, 9600, 19200, 38400 and 115200
|
Word Length: 8 bits, no parity, 1 stop bit
|
Handshake: Hardwired, modem
|
SETTLING TIME
|
From Power On: <10 s
|
ENVIRONMENTAL EFFECTS
|
Storage Temperature: –25 to +70°C (–13 to + 158°F)
|
Operating Temperature: –10 to +50°C (14 to 122°F)
|
Effect of Temperature: < 0.5 dB (–10 to +50°C)
|
Effect of Humidity: < 0.5 dB for 30%< RH< 90% (at 40°C, 1kHz)
|
BATTERIES
|
Four 1.5V LR6/AA alkaline cells
|
Lifetime (at room temperature): Typically > 10 h (with filter set selected typically > 7 h)
|
EXTERNAL DC POWER SUPPLY
|
Voltage: Regulated 7 to 15 V
|
Power: Approximately 150mA at 7V (approximately 210mA with filter set selected)
|
WEIGHT AND DIMENSIONS
|
460 g (1 lb 2 oz) (with batteries), 257 × 97 × 41mm
|
LANGUAGE
|
Each instrument is loaded with English, German, French, Italian and Spanish text. You can select one of these languages at any time
|