Type 4447 complies with the following national and international standards:
|
. ISO8041:2005: Human response to Vibration. Measuring Instrumentation
|
. ISO5349.1:2001: Mechanical Vibration. Measurement and Evaluation of Human Exposure to Hand-transmitted Vibration.
|
Part 1: General Requirements
|
. ISO5349 .2:2001: Mechanical Vibration. Measurement and Evaluation of Human Exposure to Hand-transmitted Vibration .
|
Part 2: Practical Guidance for Measurement at the Workplace
|
. ISO2631.1:1997: Mechanical Vibration and Shock. Evaluation of Human Exposure to Whole-body Vibration. Part 1: General
|
Requirements
|
. EUDirective 2002/44/EC
|
SHOCK LIMITS
|
Max. Shock level for recommended transducers (± peak): 50 km/s2
|
BATTERY
|
Rechargeable Li-ion battery 3.7 V, 2600 mA. Up to 4 h continuous use at room temperature after more than 6 h charging with the supplied charger, ZG-0459
|
Note 1: If the instrument is operated at low temperatures, the operational time is reduced
|
Note 2: External Charger (ZG-0459) is not recommended for use during measurements as it may introduce noise
|
CLOCK
|
Real-time clock and time-stamped measurements
|
INPUT CHANNELS
|
Type 4447 has 3 and 1 analogue input channels and a USB digital I/O.
|
The inputs are designed for triaxial and/or single-axis accelerometers and are equipped with selectable CCLD power supplies. Input channel sensitivity is designed for typical transducers used in human vibration measurements
|
CABLES
|
The maximum accepted accelerometer cable length is 3m (9.84 ft)
|
DETECTOR
|
Simultaneous measurement of weighted RMS, VDV, MTVV and Peak vibration value for each channel
|
FREQUENCY WEIGHTING
|
Filters for frequency weightings conform to ISO 8041:2005, including filters Wh, Wd, Wk, and Wm
|
DISPLAY
|
Color graphical display with a resolution of 124 ×124 pixels
|
Basic information regarding the instrument status is shown through icons, including:
|
Battery indicator
|
Measurement status: Measure, Pause, Stop
|
Channel status
|
Overload (O)
|
Under-range (U)
|
Elapsed time: hh:mm:ss
|
CALIBRATION
|
Calibration Check Vibration Value: 10 m/s2 (3.16m/s2 for high sensitivity accelerometers . positions 5 and 10)
|
Calibration Check Frequency: 159.2Hz
|
Electrical Calibration Check Voltage: 100mV for Type 4515-B-002 and 10 mV for Type 4520
|
PHYSICAL DIMENSIONS
|
Size: 70 × 135 × 28 mm (2.7 × 5.3 × 1.1 in)
|
Weight: 260 g (9.2 oz.), battery included
|