STANDARDS
|
Conforms with the following:
|
• IEC/EN 61672 (2002) Class 1
|
• IEC60651 Type 1 (1979) and amendment 1 (1993) and Amendment 2 (2000)
|
• IEC60804 Type 1 (2000)
|
• EN60651 Type 1 and Amendment 1 (2000)
|
• EN60804 Type 1 and Amendment 1 (2000)
|
• ANSI S1.4 –1983 Type S1
|
• ANSI S1.43 – 1997 Type 1
|
NOISE FLOOR
|
Below measurement range – less than 30 dB
|
DETECTORS
|
Simultaneous RMS and Peak with independent frequency weightings
|
Linearity Range: 70 dB
|
Pulse Range: 73 dB
|
Non-linear Distortion: insignificant
|
Peak Detector Rise Time: Typically 50 μs (< 100 μs)
|
FREQUENCY WEIGHTINGS
|
RMS: A or C
|
Peak: C
|
MICROPHONE
|
Type 4188 Prepolarized Free-field 1/2″ Condenser Microphone
|
Sensitivity: –30 dB re 1V/Pa }2dB
|
Frequency Range: 8Hz to 16 kHz }2dB
|
Capacitance: 12 pF
|
TIME WEIGHTINGS
|
F, S, I (Fast, Slow, and Impulse)
|
PARAMETERS
|
Types: Leq, MaxP, MaxL, MinL, Peak, SPL, Inst.
|
Resolution: 0.1 dB
|
Updated: Once per second
|
EXCHANGE RATE
|
3dB
|
MEMORY
|
40 Records of Measurement Results
|
CLOCK
|
Real-time (calendar) and measurement duration
|
VIBRATION SENSITIVITY
|
<80dB at 1m/s2 horizontally
|
<85dB at 1m/s2 vertically
|
AC OUTPUT
|
Short-circuit protected LEMO series 00 socket
|
Max. Output: 0.5 V RMS
|
Output Resistance: 100Ω
|
Output: Linear
|
DISPLAY
|
4 line back-lit LCD showing:
|
• Input signal level – indicated with a quasi-analog bar (updated 15
|
times per second)
|
• Selected parameter with level
|
• Warnings for overload and low battery power
|
• Measuring range
|
• Time and frequency weighting
|
• Elapsed measurement time
|
• Menus for displaying and editing settings
|
• Stored measurement results can be recalled
|
BATTERIES
|
Four 1.5V LR6/AA size alkaline cells
|
Lifetime (at room temperature): Typically >12 h
|
EFFECT OF MAGNETIC FIELD
|
80A/m (1Ørsted) at 50 Hz gives < 34 dB
|
SERIAL INTERFACE
|
Compatible with:
|
• EIA–574
|
• EIA–232–E with 25-pole adaptor
|
Baud Rate: 9600
|
Data Bits: 8
|
Stop Bit: 1
|
Parity: None
|
Handshake: XON/XOFF
|
ENVIRONMENTAL EFFECTS
|
Storage Temp.: –25 to +60°C (–13 to +140°F)
|
Operating Temp.: –10 to +50°C (14 to 122°F)
|
Temperature Effect: <0.5dB (–10 to +50°C)
|
Humidity Effect: <0.5dB for 30%<RH<90% (at 40°C, 1 kHz)
|
PHYSICAL CHARACTERISTICS
|
Size: 257×97×41 mm (10.1″×3.8″×1.6″)
|
Weight: 460 g (1.01 lb)(including batteries)
|