STANDARDS
|
Conforms with the following:
|
• IEC61672 – 1:2002 Class 1
|
• IEC60651 Type 1 (1979) with amendments 1 and 2
|
• IEC60804 Type 1 (2000)
|
• ANSIS1.4 – 1983 Type S1
|
• ANSIS1.43 – 1997 Type 1
|
PARAMETERS
|
Four parameters are measured simultaneously:
|
• LAF, instantaneous sound pressure level, frequency weighting ‘A’ and time weighting ‘F’
|
• LAFmax, maximum sound pressure level, frequency weighting ‘A’ and time weighting ‘F’
|
• LAeq, equivalent continuous sound pressure level, frequency weighting ‘A’
|
• LCpeak, maximum peak sound pressure level, frequency weighting ‘C’
|
Resolution: 0.1 dB for all four parameters
|
MEASUREMENT CONTROL
|
Measurements are manually controlled
|
Measurement times between 1 s and 60 min
|
MEASURING RANGES
|
RMS: Total range: 30 – 140 dB
|
Two manually selected ranges: 30 – 110 dB and 60 – 140 dB
|
Peak: 60 – 143 dB
|
NOISE FLOOR
|
Below measurement range, < 22 dB
|
DETECTORS
|
Simultaneous RMS and Peak with independent frequency weightings
|
Linear Operating Range: 80 dB
|
FREQUENCY RANGE
|
20Hz to 16 kHz
|
FREQUENCY WEIGHTINGS
|
A-weighting (RMS)
|
C-weighting (Peak)
|
TIME WEIGHTING
|
‘F’ (Fast)
|
OVERLOAD AND UNDER-RANGE INDICATORS
|
Overload is indicated at full-scale +0.3 dB
|
An overload detected during a timed measurement latches and is displayed with the relevant parameters until these are cleared Under-range is indicated at scale lower limit –0.5 dB
|
MICROPHONE
|
Type 4188 Prepolarized Free-field 1/2″ Condenser Microphone
|
Sensitivity: –30 dB re 1V/Pa } 2 dB (corresponding to 31.6mV/Pa)
|
Frequency Range: 8Hz to 16 kHz }2dB
|
DISPLAY
|
LCD showing:
|
• Input signal level – indicated with a quasi-analogue bar
|
• Selected parameters with level
|
• Warnings for: low battery, measurement in progress, overload and under-range
|
• Measuring range
|
• Frequency and time weighting
|
• Elapsed measurement time
|
Update Cycle: 1 s (numeric), 0.1 s (bar graph)
|
Elapsed Time Display: from 00 min 00 s up to 60 min 00 s
|
SETTLING TIME
|
Less than 5 s
|
REFERENCE CONDITIONS
|
Reference Frequency: 1000 Hz
|
Reference Sound Pressure Level: 94 dB
|
Reference Temperature: 20°C (68°F)
|
Reference RH: 65%
|
Reference Range: 30 – 110 dB
|
Reference Direction of Incidence: Frontal
|
ENVIRONMENTAL EFFECTS
|
Storage Temperature: –25 to 60°C (–13 to 140°F)
|
Operating Temperature: –10 to 50°C (14 to 122°F)
|
Effect of Magnetic Field: 80 A/m (1 oersted) at 50 Hz gives less than 30 dB (A-weighted)
|
BATTERIES
|
Two 1.5 V LR 6/AA-size alkaline batteries
|
Power Consumption During Normal Operation: Less than 300mW
|
Lifetime (at room temperature): Approximately 16 h
|
PHYSICAL CHARACTERISTICS
|
Dimensions: 230 × 78 × 31 mm including microphone
|
Weight: 245 g (8.64 oz) including batteries
|