BULK CURRENT INJECTION AS PER ISO 11452-4 |
Output level |
As required in ISO 11452-4, using closed loop or substitution method |
BULK CURRENT INJECTION AS PER MIL 461 CS 114 |
Output level |
As required in MIL 461 CS 114, using the closed loop method |
MEASUREMENTS FOR BULK CURRENT INJECTION |
Directional coupler |
Included to measure forward power and reverse power |
Forward power |
Internal power meter #1 |
Reverse power |
Internal power meter #2 |
Injected current |
Internal power meter #3 |
TEST ROUTINES FOR BULK CURRENT INJECTION |
ISO 11452-4 |
Operation via icd.control |
MIL 461 CS114 |
Operation via icd.control |
IEC 61000-4-6 |
Output level |
1V - 30Vrms (emf) all standard test levels are guaranteed with all coupling methods |
MEASUREMENTS, IEC 61000-4-6 |
Cal in (BNC) |
Integrated power meter to record the calibration data of a coupling device |
Injected current |
Measured by internal power meter |
RF indicator |
LED indicating the RF output status |
LCD |
Display of the test level and the preselected frequency value |
OUTPUT |
RF output |
N connector at the front panel |
Ouptut power |
Built-in amplifier
50dBm (9kHz - 400MHz) at 1dB compression. |
Gain amplifier |
> 50dB |
Output impedance |
50ohm |
Harmonic distortion |
< -20dBc at max. power |
Insertion loss |
Approx. 1dB (directional coupler + RF relay) |
GENERAL DATA |
Dimensions |
19""/6HU |
Weight |
approx. 31kg |
Supply voltage |
115V to 230V +10/-15%, 50/60Hz |
Input power |
Max. 380W |
Power factor |
cos (phi) = 0.96 at max. output power as per IEC 555 |
Fuses |
2 x 6.3AT (115V) or
2 x 3.15AT (230V) |
Cooling |
Active cooling, air ventilation |
Temperature |
10°C to 40°C |
Rel. humidity |
Max. 85%, non condensing |