Test voltage |
200V - 30,000V ± 5% for contact discharge
200V - 30,000V ± 5% for air discharge |
Resolution |
100V Steps |
Holding time (AD) |
More than 5s |
HV power supply |
Up to 30kV DC |
Battery |
Approx. 1-hour operation; Buffer battery is automatically recharged when connected to AC mains supply |
Interface |
USB to control the ESD generator and to download pre-programmed test routines from esd.control |
Controls |
2 function keys and one control wheel |
Display |
Graphic LCD display |
Test level |
200V to 30,000V |
Polarity |
Positive, negative; selectable |
Discharge mode |
Air or Contact discharge, selectable |
Discharge tip |
Point and ball type |
Indicators |
Integrated sensor to measure temperature and rel. humidity |
Single pulse |
Release of a single pulse |
Multiple pulses |
Continuous discharge up to 20Hz |
Counter |
0 - 9,999 for discharge pre-selection |
Contact Discharge Pulse |
As per EN/IEC 61000-4-2; 150pF/330ohm |
Rise time |
As per EN/IEC 61000-4-2 Ed.2
0.8ns +/- 25% |
First peak current value |
As per EN/IEC 61000-4-2 Ed.2
7.5A ± 15% @2kV
15A ± 15% @4kV
22.5A ± 15% @6kV
30A ± 15% @8kV |
Current value at 30ns |
As per EN/IEC 61000-4-2 Ed.2
4.0A ± 30% @2kV
8.0A ± 30% @4kV
12.0A ± 30% @6kV
16.0A ± 30% @8kV |
Current value at 60ns |
As per EN/IEC 61000-4-2 Ed.2
2.0A ± 30% @2kV
4.0A ± 30% @4kV
6.0A ± 30% @6kV
8.0A ± 30% @8kV |
Weight |
Approx. 5.1kg (ESD 30N control unit) |
Weight |
Approx. 1,250g (P 30N discharge gun) |
Supply voltage |
AC: 88V - 264V, 50/60Hz
DC: 11V - 16VDC |
Fuses |
2x 0.5A T |