Description |
Compact ESD simulator with microprocessor based, large touch-sensitive LCD panel, built-in HV relay for contact-discharge, mains operation |
Basics |
Discharge pistol, high voltage base unit, mains power supply adapter, (100 to 250 VAC), discharge network 150 pF/330 Ω, air-and contact-discharge tips, grounding cable, user manual |
Pulse data |
Standard: Conforms to IEC/EN 61000-4-2 (2001) Special: Interchangeable networks for other standards |
Pulse networks |
Network 150 pF/330 Ω as per IEC/EN 61000-4-2 (included)
Optional ISO 10605 networks 330 pF/330Ω 150 pF/2 kΩ and 330 pF/2 kΩ
Range of RC-networks for other standards:
R = 0 Ω to 20 kΩ
C = 50 to 2000 pF |
Discharge |
Air-discharge: 200 V to 30 kV (in 100 V steps)
Contact-discharge: 200 V to 30 kV (in 100 V steps) |
Discharge tips |
Ball and point as per IEC and specials; exchangeable by threaded cap |
|
Charging voltage measurement: kV, accuracy better than ±5% (stabilized) |
Discharge detection |
Air-discharge only: indicated by the kV symbol being displayed in inverse, also acoustically in the single€˜s operating mode.
Threshold level on/off. |
Holding time |
> 5 s (charging voltage ±5%) |
Charge resistor |
50 MΩ |
Triggering |
Trigger button in hand-grip |
Operation |
Via touch panel and microprocessor |
Discharge modes |
Air-discharge/contact-discharge |
Polarity |
Positive, negative/automatic change |
Operating modes |
Single/repetitive, Pulse counter 0 to 9999
Pre-select counter 0 to 9999
Repetition:
- 0.5/1/5/10/20 or 25 Hz (air)
- 0.5/1/5/10 or 20 Hz (contact)
- Continuous operation |
Auto-shut-off |
After 15 minutes idle time (without loss of the test parameters) |
Display |
LCD panel showing: charging voltage, breakdown event, polarity, air-/contact-discharge, counter/preselect counter content |
Weight |
Discharge pistol (w/o cable): 1.2 kg (2.6 lbs) approx.
Base unit: 5.7 kg (12.6 lbs) approx. |
Ambient conditions |
Operation:
+5 to +40°C
20 to 80% r.h. (non-condensing)
68 to 106 kPa |