Parameter |
Value |
Basic set |
Carrying case with: electrostatic discharge simulator, battery pack, battery charger 100 to 240 VAC, discharge network 150 pF/330 Ω, air- and contact-discharge tips, grounding cable, user manual |
Pulse
network |
Network 150 pF/330 Ω as per IEC/EN 61000-4-2 (included)
Range of R/C networks for other standards: R = 0 Ω to 10 kΩ; C = 60 to 500 pF |
Discharge voltage |
Air-discharge: 200 V to 8.8 kV (in 100 V steps)
Contact-discharge: 200 V to 6.6 kV (in 100 V steps) |
Charging voltage measurement |
kV, accuracy better than ±5% (stabilized); measurement and display of true air flashover voltage |
Discharge detection |
Indicated by an oval around the kV symbol, also acoustically in the 'single' operating mode (air-discharge only) |
Holding time |
>5 s (charging voltage ... ±5%) |
Operating modes |
Single/repetitive =
pulse counter: 0 to 9999, preselect counter 0 to 9999
Repetition: 0.5/1/5/10/20 or 25 Hz (air); 0.5/1/5 or 10 Hz (contact); continuous operation |
Displaytext |
LCD panel showing: charging voltage, discharge voltage, polarity, air-/contact-discharge, counter/preselect counter content, soft-key functions, battery monitor |
Weight |
NSG 434 with battery: 1.2 kg (2.6 lbs) approx. |
Ambient conditions |
Operation: +5 to +40°C, 20 to 80 % r.h. (non-condensing), 68 to 106 kPa |