Frequency range |
9 kHz to 2080 MHz |
Under ranging |
Down to 5 kHz (without guarantee of specs) |
Resolution |
1 Hz |
Level |
Range |
-140 to +13 dBm; -134 to +19 dBm) |
Over ranging (without guarantee on specs) |
Up to +19 dBm; -140 to +25 dBm |
Resolution |
0.1 dB |
Accuracy for levels >-127 dBm |
±1 dB (f>1.04 GHz: ±1.5 dB) |
Frequency response at 0 dBm |
1 dB, typ. 0.3 dB |
Characteristics impedance |
50 „¦ |
VSWR |
<1.5 (f>1.04 GHz: <1.8) |
Setting time (IEC/IEEE bus) |
<25 ms (<10 ms with electronic level setting) |
Non-interrupting level setting |
0 to -20 dBm |
Overload protection |
protects the instrument against externally applied RF power and DC voltage (50 W source) |
Max. permissible RF power |
30 W (SMY02: 50 W) |
Max. permissible DC voltage |
35 V |
Max. pulse load (pulse width <10 ms) |
1 mWs or 150 V (peak) |
Simultaneous modulation |
any combination of AM, FM (±·M) and pulse modulation |
Amplitude modulation |
Internal, external AC/DC |
Frequency modulation |
Internal, external AC/DC |
Phase modulation |
Internal, external AC/DC |
Pulse modulation |
external |
Internal modulation generator |
1 Hz t0 500 kHz/0.1 Hz |
RF Sweep |
digital sweep in discrete steps |
General data |
Remote control |
IEC 625 (IEEE 488) |
Power supply |
100 V/230 V (AC) -10 to +15%
120 V/220 V (AC) -12.5 to +10%
47 to 440 Hz, max. 120 VA |
Dimensions (W x H x D) |
435 mm x 147 mm x 460 mm |
Weight |
13 kg (fully equipped unit) |