| TECHNICAL DETAILS |
| Burst Module EFT/N5 as per EN/IEC 61000-4-4 and EN 61000-6-1, -6-2 |
| Test voltage |
200V - 5,500V ± 10%;
100V - 2,750V ± 10% into 50ohm |
| Pulse shape |
5/50ns into 50ohm and 1,000ohm |
| Rise time tr |
5ns ± 30% into 50ohm;
5ns ± 30% into 1,000ohm |
| Pulse width td |
50ns ± 30% into 50ohm;
50ns -15/+100ns into 1,000ohm |
| Source impedance |
50ohm |
| Polarity |
Positive/negative |
| COMBINATION WAVE/SURGE |
| Surge Module VCS/N5 as per EN/IEC 61000-4-5 and EN 61000-6-1, -6-2 |
| Voltage (o.c.) |
160V - 5,000V ± 10% |
| Pulse front time |
1.2us ± 30% |
| Pulse time to half value |
50us ± 20% |
| Current (s.c.) |
Max. 2,500A ± 10% |
| Pulse front time |
8us ± 20% |
| Pulse time to half value |
20us ± 20% |
| Polarity |
Positive/negative/alternating |
| Event counter |
1 - 30,000 or endless, selectable |
| POWER FAIL, DIPS & INTERRUPTS, VOLTAGE VARIATIONS |
| Power Fail Module PFS/N5 as per EN/IEC 61000-4-11, IEC/EN 61000-4-29 and EN 61000-6-1, -6-2 |
| Channel PF1/PF2 |
AC voltage: max. 300V
AC current: max. 16A
DC voltage: max. 300V
DC current: max. 16A |
| Frequency |
16Hz - 500Hz and DC |
| Switching time |
< 5us into a 100ohm resistive load |
| Inrush current |
> 500A |
| Protection |
Both channels are protected against short-circuit conditions. |
| GENERAL DATA |
| Interfaces |
| Serial interface |
USB |
| Parallel interface |
IEEE 488, addresses 1 - 30 |
| Analog output |
0 - 10VDC to control an external transformer |
| CN interface |
15pin SubD connector to control an external coupling network |
| Fail inputs |
DUT monitoring via Fail1 and Fail2 input (one each) |
| Dimensions |
| Dimensions |
19"", 3HU |
| Weight |
approx. 25kg |