| TECHNICAL DETAILS |
| COMBINATION WAVE 1.2/50US - 8/20US |
| Voltage (o.c.) |
500V - 10,000V ± 10% |
| Pulse front time |
1.2μs ± 30% |
| Pulse time to half value |
50μs ± 20% |
| Current (s.c.) |
Max. 5,000A ± 10% |
| Pulse front time |
8μs ± 20% |
| Pulse time to half value |
20μs ± 20% |
| Polarity |
Positive/negative/alternating |
| Event counter |
1 - 30,000 or endless |
| TRIGGER |
| Trigger of events |
Automatic, manual, external |
| CRO trigger |
5V trigger signal for oscilloscope |
| Synchronization |
0° - 360°, resolution 1° |
| Repetition rate |
6s - 999s, depending on the voltage |
| OUTPUT |
| Direct |
2 outputs via HV connectors;
- Zi = 2ohm
- for external coupling networks |
| VCS 500N10 |
10kV Surge Generator
1.2/50us - 8/20us; IEC coupling; CDN external; AC 480 V(p-p) |
| INTERFACE |
| Optical interface |
Opto link, 3m cable
USB A connector |
| Parallel serial interface |
IEEE 488, addresses 1 - 30 |
| CN interface |
To control external coupling matrix |
| SAFETY |
| Safety circuit |
Control input (24Vdc) |
| Warning lamp |
Floating output contact |
| GENERAL DATA |
| Dimensions - Weight |
19""/6HU, approx. 35kg |
| Supply voltage |
115/230V +10/-15% |
| Fuses |
2x T 2AT (230V) or 2X T 4AT (115V) |