| TECHNICAL DETAILS |
| VCS 500N12 |
12 kV Surge Generator
1.2/50 us-8/20 us, IEC and ANSI Cat.B coupling, with ext. coupler 3x480 V |
| COMBINATION WAVE 1.2/50US - 8/20US |
| Voltage (o.c.) |
500V - 12,000V ±10% |
| Rise time |
1.2us ± 30% |
| Pulse time to half value |
50us ± 20% |
| Current (s.c.) |
max. 6,000A ±10% |
| Rise time |
8us ± 20% |
| Pulse time to half value |
20us ± 20% |
| Polarity |
Positive/negative/alternating |
| Event counter |
1 - 30,000 or endless |
| PULSE OUTPUT |
| Direct |
Via HV-connectors:
- Zi = 2 ohm
- To connect external surge couplers |
| INTERFACE |
| Optical interface |
Opto link, 3m cable
USB A connector |
| Parallel interface |
IEEE 488, addresses 1 - 30 |
| SAFETY |
| Safety circuit |
Control input (24 Vdc) |
| Warning lamp |
Floating output contact |
| CN interface |
To control external coupling matrix |
| GENERAL DATA |
| VCS 500N12 |
| Dimensions |
19""/9 HU |
| Weight |
approx. 35kg |
| SUPPLY |
| Supply voltage |
115/230V +10/-15% |
| Fuses |
2x2 AT (230V) or 2x4 AT (115V) |