NETWAVE 30.3 |
Output voltage |
0 V - 3*400 V AC (p-n)
0 V - 3*690 V AC (p-p)
0 V - +/- 1120 V DC |
Output current (@ max. 300 V AC/360V DC) |
33 A (RMS) continuous
66 A (RMS) short term (max. 3 s)
250 A repetitive peak |
Power Recovery 30 |
Included |
GENERAL DATA (ALL MODELS) |
SPECIFICATIONS |
Output frequency |
DC - 5,000 Hz |
Frequency accuracy |
100 ppm |
Output connectors |
Safety lab connectors
CEE type 32 A (only for NetWave 20.x and NetWave 30.x) |
Interfaces |
GPIB
Ethernet
RS 232 (input from DPA analyser)
Frame bus (internal system bus) |
Phase accuracy |
Resolution 1° |
Output noise |
< 50 V: 110 mV rms
> 50 V: 320 mV rms |
Slew Rate |
8V/μs |
REGULATION |
Voltage sense |
Internal or external, 4 wires |
Distortion (THD) |
Less than 0. 5%, @50/60 Hz |
Output voltage Stability |
Better than 0.1 % |
Output voltage Accuracy |
Better than 0.5 % |
Max. compensable drop on wires |
5 % of V nominal |
Current limiter |
5 A to Imax
Stop / Continuous (behavior) |
Protection |
Over current
Over voltage
Over temperature
Low voltage |
TRIGGER AND DUT MONITORING |
Trigger |
2 inputs, 2 outputs |
DUT monitors |
2 inputs, configurable |
OPTION FOR DC EXTENDED VOLTAGE RANGE |
* with option |
OPT-3 DC-EVR,
Double the DC voltage |