| TECHNICAL DETAILS | 
| COMBINATION WAVE 1.2/50US - 8/20US | 
| Voltage (o.c.) | 500V - 10,000V ± 10% | 
| Pulse front time | 1.2μs ± 30% | 
| Pulse time to half value | 50μs ± 20% | 
| Current (s.c.) | Max. 5,000A ± 10% | 
| Pulse front time | 8μs ± 20% | 
| Pulse time to half value | 20μs ± 20% | 
| Polarity | Positive/negative/alternating | 
| Event counter | 1 - 30,000 or endless | 
| TRIGGER | 
| Trigger of events | Automatic, manual, external | 
| CRO trigger | 5V trigger signal for oscilloscope | 
| Synchronization | 0° - 360°, resolution 1° | 
| Repetition rate | 6s - 999s, depending on the voltage | 
| OUTPUT | 
| Direct | 2 outputs via HV connectors;
- Zi = 2ohm
- for external coupling networks | 
| VCS 500N10.3 | 10kV Surge Generator
1.2/50us - 8/20us, IEC & ANSI Cat.B coupling; CDN external; AC 480 V(p-p);
6HU; used with CNV 503N10.1 | 
| INTERFACE | 
| Optical interface | Opto link, 3m cable
USB A connector | 
| Parallel serial interface | IEEE 488, addresses 1 - 30 | 
| CN interface | To control external coupling matrix | 
| SAFETY | 
| Safety circuit | Control input (24Vdc) | 
| Warning lamp | Floating output contact | 
| GENERAL DATA | 
| Dimensions - Weight | 19""/6HU, approx. 35kg | 
| Supply voltage | 115/230V +10/-15% | 
| Fuses | 2x T 2AT (230V) or 2X T 4AT (115V) |