| Frequency Response (w/calibration factors applied) |
±1.8 dB, 3MHz-10 MHz |
| ±0.5 dB, 10MHz-200MHz |
| ±1.4 dB, 200MHz-8 GHz |
| ±1.8 dB, 8.0 GHz-18 GHz) |
| Analog Response Time |
50 usec |
| Isotropic Deviation |
± 1.5 dB (10 MHz-8 GHz) |
| ± 2 dB (f >8 GHz) |
| Sensitivity |
0.6-1000 V/m |
| Linearity |
± 3 dB (0.6-1.65 V/m) |
| ± 1 dB (1.65-3.3 V/m) |
| ± 0.5 dB (3.3-300 V/m) |
| ± 0.7 dB (300-1000 V/m) |
| Damage Level |
1600 V/m CW and 16,000 V/m peak |
| Probe Sampling Rate |
60Hz, 50Hz, 5 Hz- user selectable |
| Data Transfer Rate |
>30 samples per second |
| Power Requirements |
2 built-in NiMH batteries, size AA, rechargeable, changed by operator, 40 hours operating life per charge. |
| Battery Charger |
Standard recharger included, 3 hour charge time. |
| Probe Head Diameter |
65mm |
| Weight |
350g approximately |
| Ranges |
Single continuous range |
| Operating Temperature Range |
0 to 50°C |
| Approximate Dimensions |
420 x 85 x 65 mm including base and probe head. |