Amplitude Accuracy (field aligned with sensor axes) |
Without correction factors applied |
±1.0 dB @ 10 MHz |
With correction factors applied |
Typical expanded measurement uncertainty (95% confidence interval) |
0.8 dB, 100 kHz-1 GHz |
1.4 dB, 1 GHz-6 GHz |
Response Time/Sampling Rate (through FI7000) |
20 msec/up to 50 samples per second, USB and GPIB only |
Isotropic Deviation (measured at the critical angle) |
± 0.5 dB @ 10 MHz |
± 0.5 dB, 0.5 MHz-2 GHz (typical) |
Operating Range |
0.5-800 V/m, 100 kHz-1 GHz |
0.5-600 V/m, 1 GHz-4 GHz |
0.7-800 V/m, 4 GHz-6 GHz |
Linearity, 0.5 to 800 V/m |
± 0.5 dB and ± 0.3 V/m |
Temperature Stability |
± 0.5 dB over operating temperature range |
Damage Level |
1000 V/m continuous field |
Ranges |
Single range |
Data returned from probe |
X, Y, Z axes, and composite |
Power Requirements |
Laser powered from FI7000 interface |
Dimensions |
5.7 x 5.7 x 5.7 cm (2.25 x 2.25 x 2.25 in) |
2.92 cm (1.15 in) DIA spherical housing |
3.18 cm (1.25 in) sensor radome per axes |
Weight |
62.5 g (2.2 oz) |
Operating Temperature Range |
10ËšC to 40ËšC (50ËšF to 104ËšF) @ 5% to 95% RH non-condensing |
Fiber Optic Connectors |
Two E2000 compact duplex connectors at 1 meter, includes fiber optic verification loop. |