| H-field probe (9 kHz to 30 MHz) |
| Frequency range |
9 kHz to 30 MHz |
| Usable frequency range |
9 kHz to 100 MHz |
| Maximum permissible voltage of uninsulated conductor |
0 Hz to 120 Hz 500 V (Vp) |
| Maximum input power EMS testing |
0.5 W |
| RF connector |
SMA female |
| Dimensions |
| W × H × D (including RF connector) |
256 mm × 38 mm × 18 mm
(10.1 in × 1.5 in × 0.7 in) |
| H-field probe (30 MHz to 1 GHz) |
| Frequency range |
30 MHz to 1 GHz |
| Usable frequency range |
1 MHz to 2 GHz |
| Maximum permissible voltage of uninsulated conductor |
0 Hz to 120 Hz 500 V (Vp) |
| Maximum input power EMS testing |
0.25 W |
| RF connector |
SMA female |
| Dimensions |
| W × H × D (including RF connector) |
256 mm × 38 mm × 18 mm
(10.1 in × 1.5 in × 0.7 in) |
| E-field probe (9 kHz to 1 GHz) |
| Frequency range |
9 kHz to 1 GHz |
| Frequency response |
±3 dB |
| Antenna factor with capacitive coupling |
13 mV/V |
| Antenna factor |
typ. 67 dB (1/m) |
| Maximum permissible voltage at probe tip |
20 V |
| RF connector |
SMA female |
| Dimensions |
| W × H × D (including RF connector) |
267 mm × 38 mm × 18 mm
(10.5 in × 1.5 in × 0.7 in) |
| DC supply |
| Required DC voltage with E-field probe |
10/15 V ± 0.1 V |
| DC connector |
LEMO (2 contacts with screen) |
| RF connector input |
BNC female |
| RF connector output |
N male |
| Dimensions |
| W × H × D (including RF connector) |
97 mm × 26 mm × 28 mm
(3.8 in ×1.0 in × 1.1 in) |