Parameter |
Value |
Basic set |
Carrying case with: discharge pistol, cradle for discharge pistol, high voltage base unit with built-in battery pack, mains adapter and battery charging unit (100 to 250 VAC), discharge network 150 pF/330 Ω, air- and contact-discharge tips, grounding cable, user manual, 25 mm discharge sphere |
Pulse networks |
Network 150 pF/330 Ω as per IEC/EN 61000-4-2 (included)
Optional ISO 10605 networks 330 pF/330Ω, 150 pF/2 kΩ and
330 pF/2 kΩ
Range of RC networks for other standards:
R = 0 Ω to 20 kΩ; C = 50 to 2000 pF |
Discharge voltage |
Air-discharge: 200 V to 30 kV (in 100 V steps)
Contact-discharge: 200 V to 30 kV (in 100 V steps) |
Discharge tips |
Ball and point as per IEC and specials, exchangeable by threaded cap |
Charging voltage measurement |
kV, accuracy better than ±5% (stabilized) |
Discharge detection (air-discharge only) |
Indicated by the kV symbol being displayed in inverse also acoustically in the €˜Single€™ operating mode. Adjustable threshold level. |