OUTPUT CHARACTERISTICS CONTINUOUS MODE |
Test frequencies |
DC, 16 2/3Hz, 50Hz and 60Hz |
Signal level |
0.1 - 35 Vrms or DC |
Test level 1 |
1V continuous |
Test level 2 |
3V continuous |
Test level 3 |
10V continuous |
Test level 4 |
30V continuous |
Output impedance |
50ohm ± 10% |
Total harmonics distortion |
< 10% (sinusoidal waveform) |
Ripple on DC |
< 5% |
OUTPUT CHARACTERISTICS SHORT-TERM MODE |
Test frequencies |
DC, 16 2/3Hz, 50Hz and 60Hz |
Signal level |
1 - 330 Vrms or DC |
Test level 1 |
10V for 1s duration |
Test level 2 |
30V for 1s duration |
Test level 3 |
100V for 1s duration |
Test level 4 |
300V for 1s duration |
Output impedance |
50ohm ± 10% |
Total harmonics distortion |
< 10% (sinusoidal waveform) |
Phase synchronization |
0° ± 5% |
Ripple on DC |
< 5% |
Fall/rise time |
between 1us and 5us |
Note: |
These tests require either the optional MV 2606N2.2 motor variac or the optional programmable AC voltage source ACS 500N2.3. |
TESTING AS PER IEC/EN 61000-4-16 OUTPUT CHARACTERISTICS 15Hz - 165kHz |
Frequency range |
10Hz to 165kHz |
Signal level |
0.1 to 35 Vrms |
Test level 1 |
1V - 0.1V - 1V |
Test level 2 |
3V - 0.3V - 3V |
Test level 3 |
10V - 1V - 10V |
Test level 4 |
30V - 3V - 30V |
Output impedance |
50ohm ± 10% |
Total harmonics distortion |
< 1% (sinusoidal waveform |
INTERFACE |
Serial interface |
USB |
Parallel interface |
IEEE 488, addresses 1 to 30 |
Fail 1 |
BNC input; test will be stopped immediately when input becomes active low) |
Fail 2 |
BNC input; test will be stopped (when input becomes active low) and continued (input active high). After 10 fail events the test will be stopped. |
GENERAL DATA |
Dimensions |
19""/6HU 555mm x 448mm x 286mm |
Weight |
Approx. 30kg |
Supply voltage |
115V - 230V +10/-15%, 50/60Hz |
Input power |
Max. 600W |
Fuses |
2 x 6.3AT (115V) or 2 x 3.15AT (230V)
2 x 10AT for AC-Source |
Cooling |
Active cooling, air ventilation |
Temperature |
10°C to 40°C |