| TECHNICAL DETAILS |
| As per EN/IEC 61000-4-4 and EN 61000-6-1, -6-2 |
| Test voltage |
200V - 5,500V ± 10%;
100V - 2,750V ± 10% into 50ohm |
| Pulse shape |
5/50ns into 50ohm and 1,000ohm |
| Rise time tr |
5ns ± 30% into 50ohm;
5ns ± 30% into 1,000ohm |
| Pulse width td |
50ns ± 30% into 50ohm;
50ns -15/+100ns into 1,000ohm |
| Source impedance |
50ohm |
| Polarity |
Positive/negative |
| Output |
|
| Direct |
Via 50ohm coaxial connector |
| Coupling mode |
L, N, PE; all combinations |
| DUT supply |
AC: 300V/16A; 50/60Hz |
|
DC: 300V/16A |
| CRO trigger |
5V trigger signal for oscilloscope |
| Interface |
| Serial interface |
USB |
| Parallel interface |
IEEE 488, addresses 1 - 30 |
| Analog output |
0 - 10VDC to control an external transformer |
| CN interface |
15pin SubD connector to control an external coupling network |
| Fail inputs |
DUT monitoring via Fail1 and Fail2 input (one each) |
| Dimensions |
| Dimensions |
19"", 3HU |
| Weight |
approx. 14kg |